| KNM-04 NIBP module

Technical Specification
Dimension applied front page
68 x 128x 32mm (W x L x H)
Weight approximately 180g
Supply Power Voltage DC 7Volt 10%, 500mA (DC5 also available)
Consumption Power 3.2 Watt at Operating Motor
0.35 Watt at Standby
Measuring Range 50 - 280mmHg Systolic, Diastolic and Mean
30 ~ 300bpm Pulse Rate
Measuring Mode Adult/Neonate
Manual/Automatic
Protection Double Overpressure Protection (Meet IEC Regulation)
Overpressure 300mmHg + 20mmHg at Adult Mode
150mmHg + 15mmHg at Neonate Mode
Hardware Protection - Use of 2 Pressure Sensor
Use of 2 Solenoid Valve
Source- Sink Driver
Software Protection
- Detect Measuring Over Time
- Detect Malfunction of Controller
Manometer Mode 0 - 300mmHg 3mmHg
Operating Temperature 10 ~ 40
Operating Humidity 0% ~ 85%
Storage Temperature -5 ~ 50
Measuring Method
Oscillometric Method
Non-invasive Pneumatic Cuff
Module and Interface
Communication
l RS-232C (MAX232) Level and UART
l Baud-Rate 9600bps
l 8 Data bit
l 1 Stop bit
|